2.
Failure
to calibrate equipment in accordance with an established written
program. Failure to document or ensure that instruments are
calibrated: transfilling pressure gauges, vacuum gauge, thermometer,
Oxygen Analyzer. (21 CFR 211.160)
3.
Failure
to ensure that each person engaged in the manufacture, processing
and transfilling of the drug product, and each person responsible
for supervising these activities, has the education, training,
and experience to enable that person to perform their assigned
functions in such a manner as to provide assurance that drug
product has the quality and purity that it purports or is
represented to possess. (21 CFR 211.25)
4.
A
designated Quality Control Unit has not been established,
or quality control unit is not performing duties. Duties such
as approve or reject all components, drug product containers,
closures, in-process materials, packaging material, labeling,
and drug products, and no review of production records to
assure that no errors had occurred or, if errors had occurred,
that they had been fully investigated. (21 CFR 211.22)
5.
Failure
to maintain appropriate batch production and control records
for each batch of drug product produced, including documentation
that each significant step in the manufacture, processing,
packing, or holding of the batch was accomplished and the
time of performance. (21 CFR 21.188)
6.
Failure
to establish written procedures designed to assure that correct
labels and labeling are used for the product cylinders, including
identification of the drug product with a lot or control number
that permits determination of the history of the manufacture
and control of the batch. (21 CFR 211.130)
7.
Failure
to perform adequate prefill operations on each product cylinder.
(21 CFR 211.84)
8.
Failure
to test or verify the results of each lot of bulk oxygen to
determine conformance with appropriate specification for identity
and strength.
9.
Failure
to establish written procedures for the handling of any complaints
regarding the drug product; and to adequately investigate
complaints and document the findings of an investigation and
follow-up. (21 CFR 211.198)
10.
Failure
to reconcile the quantities of medical gas labels issued and
returned is not performed. (21 CFR 211.125(c))
11.
Standard
Operating Procedures do not contain information concerning
the generation of tank lot numbers, or there are no Standard
Operating Procedures.
12.
Failure
to establish detailed written procedures to cover the various
aspects of transfilling operation. Batch production and control
records for cryogenic vessels are not maintained to show that
each significant step in the manufacturing operation was accomplished.
13.
Failure
to establish separate quarantine areas for the holding of
empty and filled cylinders of Oxygen USP. Failure to have
designated quarantine areas for empty cylinders, rejected
cylinders, and finished medical gas product cylinders prior
to release. (21 CFR 211.42)
14.
Failure
to establish written procedures with respect to the testing
of Oxygen USP; the calibration of the oxygen analyzer; and
the documentation of prefill and postfill inspections of product
cylinder home units. (21 CFR 211.100(a))
15.
Failure
to store medical gas labels in an area limited to authorized
personnel. (21 CFR 211.122(d))
16.
Failure
to complete employee training, or inadequate training, in
GMP’s.
17.
Failure
to retain production, control or distribution records associated
with Oxygen, USP for at least 1 year after the expiration
date of the batch or, for products lacking expiration dating,
a minimum of three years after distribution of the batch.
Failure to take adequate action to correct deficiencies identified
during internal audits. (21 CFR 211.180)
18.
Failure
to establish written procedures to determine the expiration
date of a drug product. (21 CFR 211.137(a))
19.
Failure
to have a master label on file in the master production and
control records. (21 CFR 211.186(b))
20.
Failure
to maintain records of the qualifications for consultants
advising on the manufacture, processing, packing or holding
of drug products. (21 CFR 211.34)
21.
Failure
to have a written record of maintenance for the oxygen analyzer.
(21 CFR 211.182)
22.
Failure
to routinely calibrate in accordance with written procedures,
vacuum gauges, pressure gauges, and thermometers. (21 CFR
211.68(a))